LOW-TEMPERATURE STERILIZATION OF HEAT-SENSITIVE MEDICAL DEVICES: A SYSTEMATIC REVIEW OF MICROBIAL EFFICACY AND MATERIAL COMPATIBILITY
Abstract
Background: Low-temperature sterilization provides processing options for medical devices that cannot tolerate conventional steam sterilization. Its suitability depends on achieving microbial inactivation while preserving material integrity and device function.
Objective: To evaluate the microbial efficacy and material compatibility of low-temperature sterilization methods for heat-sensitive medical devices, considering device characteristics, processing conditions, and repeated exposure.
Methods: A systematic review is planned using MEDLINE, Embase, Scopus, and Web of Science, supplemented by citation searching. Eligible primary studies will report microbial efficacy, material compatibility, or both. Two reviewers will independently assess eligibility and methodological limitations. Findings will undergo structured narrative synthesis, with meta-analysis considered where studies are sufficiently comparable. The present abstract summarizes a preliminary synthesis of four verified primary studies; systematic screening and formal appraisal remain outstanding.
Preliminary Results: Comparative laboratory studies reported differences in microbial inactivation across sterilization systems under specified contamination and device-challenge conditions. A study of selected three-dimensional printed materials supported vaporized hydrogen peroxide sterilization within defined material and design limits. A separate investigation of seven epoxy formulations exposed to 100 vaporized hydrogen peroxide cycles classified six as compatible and one as incompatible because of changes in texture and hardness. Differences in experimental conditions and outcomes prevented a defensible overall ranking of technologies. No pooled estimates were calculated.
Conclusion: The preliminary evidence supports evaluating specific device–process combinations using both microbial and compatibility outcomes. Favorable findings in either domain alone cannot establish overall suitability. Definitive comparative conclusions require completion of the systematic search, eligibility assessment, and methodological appraisal.
How to Cite
References
- 1. Centers for Disease Control and Prevention. (2023a, November 28). Ethylene oxide “gas” sterilization. https://www.cdc.gov/infection-control/hcp/disinfection-sterilization/ethylene-oxide-sterilization.html
- 2. Centers for Disease Control and Prevention. (2023b, November 28). Low-temperature sterilization technologies. https://www.cdc.gov/infection-control/hcp/disinfection-sterilization/low-temperature-sterilization.html
- 3. Centers for Disease Control and Prevention. (2023c, December 7). Recommendations for disinfection and sterilization in healthcare facilities.
- https://www.cdc.gov/infection-control/hcp/disinfection-sterilization/summary-recommendations.html
- 4. Eveland, R., Antloga, K., Meyer, A., & Tuscano, L. (2024). Low temperature vaporized hydrogen peroxide sterilization of 3D printed devices. 3D Printing in Medicine, 10, Article 6. https://doi.org/10.1186/s41205-024-00206-1
- 5. Gitsov, I. P. I., Liu, Y., Saylor, D. M., Hendrickx, A. L., Gronda, A. M., Sinha, M. N., Long, E., Liu, S., Hu, R., Weeks, J. W., Eppihimer, M., & Chaffin, K. A. (2025). Diffusion of hydrogen peroxide through medical grade poly(ether)urethane: Analyzing mechanisms of sorption and transport to support sterilization with vapor phase hydrogen peroxide. Journal of Biomedical Materials Research Part B: Applied Biomaterials, 113(5), Article e35567. https://doi.org/10.1002/jbm.b.35567
- 6. International Organization for Standardization. (2014). Sterilization of health-care products—Ethylene oxide—Requirements for the development, validation and routine control of a sterilization process for medical devices (ISO Standard No. 11135:2014). https://www.iso.org/standard/56137.html
- 7. International Organization for Standardization. (2022). Sterilization of health care products—Low temperature vaporized hydrogen peroxide—Requirements for the development, validation and routine control of a sterilization process for medical devices (ISO Standard No. 22441:2022).
- https://www.iso.org/standard/73214.html
- 8. Kanemitsu, K., Imasaka, T., Ishikawa, S., Kunishima, H., Harigae, H., Ueno, K., Takemura, H., Hirayama, Y., & Kaku, M. (2005). A comparative study of ethylene oxide gas, hydrogen peroxide gas plasma, and low-temperature steam formaldehyde sterilization. Infection Control & Hospital Epidemiology, 26(5), 486–489. https://doi.org/10.1086/502572
- 9. McEvoy, B., & Rowan, N. J. (2019). Terminal sterilization of medical devices using vaporized hydrogen peroxide: A review of current methods and emerging opportunities. Journal of Applied Microbiology, 127(5), 1403–1420. https://doi.org/10.1111/jam.14412
- 10. Padhee, S., Fingar, C., Patel, A., Eveland, R., Rantucci, J., Ward, T., Nandivada, V., & Ramnath, R. (2024). Material compatibility of epoxies exposed to repeated low temperature vaporized hydrogen peroxide sterilization. Polymer-Plastics Technology and Materials, 63(18), 2511–2520.
- https://doi.org/10.1080/25740881.2024.2376209
- 11. Page, M. J., McKenzie, J. E., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., Shamseer, L., Tetzlaff, J. M., Akl, E. A., Brennan, S. E., Chou, R., Glanville, J., Grimshaw, J. M., Hróbjartsson, A., Lalu, M. M., Li, T., Loder, E. W., Mayo-Wilson, E., McDonald, S., . . . Moher, D. (2021). The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ, 372, Article n71. https://doi.org/10.1136/bmj.n71
- 12. Rutala, W. A., Gergen, M. F., & Weber, D. J. (1998). Comparative evaluation of the sporicidal activity of new low-temperature sterilization technologies: Ethylene oxide, 2 plasma sterilization systems, and liquid peracetic acid. American Journal of Infection Control, 26(4), 393–398. https://doi.org/10.1016/S0196-6553(98)70034-3
- 13. Rutala, W. A., Gergen, M. F., Sickbert-Bennett, E. E., & Weber, D. J. (2020). Comparative evaluation of the microbicidal activity of low-temperature sterilization technologies to steam sterilization. Infection Control & Hospital Epidemiology, 41(4), 391–395. https://doi.org/10.1017/ice.2020.2
- 14. U.S. Food and Drug Administration. (2024, January 8). FDA facilitates broader adoption of vaporized hydrogen peroxide for medical device sterilization. https://www.fda.gov/news-events/press-announcements/fda-facilitates-broader-adoption-vaporized-hydrogen-peroxide-medical-device-sterilization